| Model Number |
FGA15090-040 |
| Medical Device Problem Code |
Difficult or Delayed Positioning (1157)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/31/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lot number was provided.The manufacturing records were reviewed.All inspection values were within specification and there were no ncr's or deviations.The device was not returned for evaluation; therefore, an analysis could not be conducted.The root cause cannot be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that a web device was successfully detached within an aneurysm, but the pusher wire was advanced forward, invaginating the web device and pushing it closer to the dome of the aneuysm.A snare was used to pull the web back into place.There was no reported adverse clinical sequela to the patient.
|
| |
|
Search Alerts/Recalls
|