• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV (USD) SHILEY; TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) SHILEY; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 86232
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the device had a sticky residue at the end of the tube.It was stated that it was cloudy and had a rough surface all throughout.There was no patient involvement.
 
Additional Manufacturer Narrative
Evaluation summary: one sample of cuffless endotracheal tube was received for evaluation.The sample was received sealed inside its pouch.A visual inspection was performed and it was observed a sticky residue on the length tube, the reported failure mode is confirmed.Corrective actions have been implemented to mitigate this issue.Information has been added to the database and trends will continue to be monitored.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
MDR Report Key8953646
Report Number2936999-2019-00645
Device Sequence Number2399909
Product Code BTR
UDI-Device Identifier10884521620230
UDI-Public10884521620230
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K151381
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/20/2022
Device Model Number86232
Device Catalogue Number86232
Device Lot Number17D0455JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2019
Initial Date Received by Manufacturer 08/08/2019
Supplement Date Received by Manufacturer10/01/2019
Initial Report FDA Received Date09/02/2019
Supplement Report FDA Received Date10/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-