Model Number 179308
Medical Device Problem Code Packaging Problem (3007)
Health Effect - Clinical Code Tissue Damage (2104)
Date of Event 08/05/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, at the beginning of the operation of a thoraco lobectomy, when inserting the trocar as the 2nd port, a crack was found at the tip of the trocar when observing with a camera.It was stated that there was a scratch like an agnail.When the trocar was removed to be replaced by a new one, the tissue was damaged as the agnail on the trocar rubbed into the tissue.The surgical time was extended by less than 30 minutes.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted: the distal end of the cannula was cracked.The obturator appeared intact.A functional evaluation found that the obturator was inserted into the cannula and removed, no resistance was noted.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the cracked cannula may occur when excessive force is applied during a clinical application.The root cause of the observed damage was found to be due to the device not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THORACOPORT
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key8954529
Report Number2647580-2019-04403
Device Sequence Number11393198
Product Code GCJ
UDI-Device Identifier10884521081215
UDI-Public10884521081215
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K992288
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model Number179308
Device Catalogue Number179308
Device Lot NumberP8H1526X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2019
Initial Date Received by Manufacturer 08/09/2019
Supplement Date Received by Manufacturer11/13/2019
Initial Report FDA Received Date09/03/2019
Supplement Report FDA Received Date11/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-