BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 24700
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/22/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The target lesion was located in a liver.A 4.0mmx40mmx135cm sterling balloon catheter was advanced for dilatation.However, the balloon failed to cross the lesion.It was then noted that during first inflation at 6 atmospheres, the head end of the balloon ruptured.The procedure was completed with a different device.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key8957705
Report Number2134265-2019-10581
Device Sequence Number2151668
Product Code DQY
UDI-Device Identifier08714729123576
UDI-Public08714729123576
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K053116
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2021
Device Model Number24700
Device Catalogue Number24700
Device Lot Number0022453434
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2019
Initial Report FDA Received Date09/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
Patient Weight48
-
-