• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number INFANT FLOW SIPAP
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/03/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Any additional information provided by the customer will be included in a follow up report.At this time, vyaire has not received the suspect device/component for evaluation.
 
Event or Problem Description
It was reported to vyaire medical that the sipap device experienced loss of pressure while connected to a patient.The device also alarmed error codes 33 and 42.The customer confirmed that there was no patient harm associated with the reported event.
 
Additional Manufacturer Narrative
Results of investigation: a vyaire failure analysis lab technician was unable to verify the customer's reported issue.The device passed all testing and met all vyaire manufacturer specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VYAIRE MEDICAL
22745 savi ranch pkwy
yorba linda CA 92887
MDR Report Key8957903
Report Number2021710-2019-10748
Device Sequence Number16774491
Product Code CBK
UDI-Device Identifier10846446005387
UDI-Public(01)10846446005387(11)20181227
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K031745
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberINFANT FLOW SIPAP
Device Catalogue Number27444-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2019
Initial Date Received by Manufacturer 08/07/2019
Supplement Date Received by Manufacturer11/11/2019
Initial Report FDA Received Date09/03/2019
Supplement Report FDA Received Date12/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-