The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The threads on both ends of the device are damaged, rendering them inoperable.The hex end is also heavily damaged from impacts.The device was manufactured in 2004 and exhibits signs of extensive wear/usage.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|