MAQUET CARDIOPULMONARY GMBH HLM TUBING SET W/BIOLINE COATING; TUBING, PUMP, CARDIOPULMONARY BYPASS Back to Search Results
Model Number BEQ-HLS 5050 USA#HLS SET ADVANCED 5.0
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Maquet cardiopulmonary (b)(4) requested the product for investigation but the product was not available.Therefore no laboratory investigation could be performed by the manufacturer.A review for similar complaints to be investigated already was performed and no similar complaints were found.Thus the reported failure could not be confirmed.Based on the information available at this time the cause of this failure was determined to not be attributed to a device related malfunction.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.The devices were not returned for investigation.
 
Event Description
They stated that they spoke with (b)(6), ccp, from (b)(6) regarding two separate hls sets in which they had oxygenation problems in the previous 2 weeks.In both cases, they had their sweep at 11 lpm and the arterial p02 were only around 65-70 at 100% fi02.They tried another 02 source and this was not the issue.In both cases, once they changed out the hls sets it fixed the issue and they were sweeping around 3 lpm.Clot was not detected nor was the delta p affected in either case.Additional information 2019-08-28: they did not keep the kits for inspection.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLM TUBING SET W/BIOLINE COATING
Type of Device
TUBING, PUMP, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
NURSEL BOELENS
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
GM   76437
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
4972229321
MDR Report Key8958965
MDR Text Key182553162
Report Number8010762-2019-00274
Device Sequence Number1
Product Code DWE
Combination Product (y/n)N
PMA/PMN Number
K080592
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/04/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/23/2020
Device Model NumberBEQ-HLS 5050 USA#HLS SET ADVANCED 5.0
Device Catalogue Number701052797
Device Lot Number70128132
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/13/2019
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/24/2018
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
-
-