Model Number UH1-54-28
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/02/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Black staining of uhr head.Update august 15th; spoke to rep.This was a primary surgery case further to trauma for a hip fracture where they were implanting exeter stem and uh1 head into the patient.Black staining was noticed and not used.Another was used instead.No impact to patient.There was no revision for this patient.
 
Event or Problem Description
Black staining of uhr head.Update august 15th; spoke to rep.This was a primary surgery case further to trauma for a hip fracture where they were implanting exeter stem and uh1 head into the patient.Black staining was noticed and not used.Another was used instead.No impact to patient.There was no revision for this patient.
 
Additional Manufacturer Narrative
An event regarding foreign matter involving uhr head was reported.The event was confirmed through visual inspection method & results: device evaluation and results: visual inspection visual inspection was performed as part of the material evaluation and indicated the following comments: device was returned without the outer foil lidding.Inner lidding has been taped closed with masking tape.Two specs of debris were located on the polymer component of the device.Dimensional inspection: not performed as the dimensional aspects are not in question.Functional inspection: not performed as the functional aspects are not in question.Material analysis: mar not performed as the reported event relates to a packaging issue.Unable to determine source of debris due to insufficient debris observed at the current time.Clinician review: no medical records were received for review with a clinical consultant.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: no further investigation for this event is possible at this time as insufficient information was received by stryker orthopaedics.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UHR BIPOLAR 28X54MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8959554
Report Number0002249697-2019-03071
Device Sequence Number3146912
Product Code JDI
UDI-Device Identifier07613327017250
UDI-Public07613327017250
Combination Product (Y/N)N
Initial Reporter CountrySE
PMA/510(K) Number
K173499
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 02/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2024
Device Model NumberUH1-54-28
Device Catalogue NumberUH1-54-28
Device Lot NumberVX5764
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/08/2019
Supplement Date Received by Manufacturer01/23/2020
Initial Report FDA Received Date09/04/2019
Supplement Report FDA Received Date02/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-