| Model Number |
7214 |
| Medical Device Problem Codes |
Break (1069); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/01/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: date the incident occurred was estimate as 01-aug-2019 which is the first day of the month of the aware date.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break and removal difficulties occurred.A 2.50mm x 20mm nc emerge balloon catheter was advanced for dilation.However, during inflation at 24 atmospheres, the shaft broke.Removal difficulty was noted so an optical coherence tomography was performed to assess the situation.Finally, the wire and the balloon were completely removed by snaring.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
B3 - date the incident occurred was estimate as (b)(6) 2019 which is the first day of the month of the aware date.Device evaluated by mfr.: returned product consisted of an nc emerge balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 108cm distal of the strain relief at the midshaft bond.There was contrast and blood in the inflation lumen and contrast in the balloon.The balloon was loosely folded.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break and removal difficulties occurred.A 2.50mm x 20mm nc emerge balloon catheter was advanced for dilation.However, during inflation at 24 atmospheres, the shaft broke.Removal difficulty was noted so an optical coherence tomography was performed to assess the situation.Finally, the wire and the balloon were completely removed by snaring.No patient complications were reported.
|
| |
|