| Model Number |
312 |
| Medical Device Problem Code |
Electrical Shorting (2926)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/29/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Corindus service personnel investigated the problem and found that a burnt cable assembly and a defective arm controller board.Both parts were replaced and the system is functioning normally.A source of the failure could not be confirmed as the user had discarded the connection cable prior to the service personnel's arrival.However, it is likely that the lemo connector became loose and lead to a damaged cable and shorted arm board.
|
| |
|
Event or Problem Description
|
|
During preparation for a hospital training event, it was observed that the robotic drive was functioning as if there was a loose connection and unexpectedly shut down.Upon investigation it was observed that a burnt port was present an electrical connector of the extended reach arm.
|
| |
|