• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number D-AVHD-DF16
Device Problem Material Frayed (1262)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/24/2019
Event Type  malfunction  
Manufacturer Narrative
One advisor¿ hd grid mapping catheter, sensor enabled¿ was received for investigation.Electrodes 12-16 met specifications for acceptable resistance values with no open or short circuits detected.Microscopic inspection revealed the distal coupler was displaced, and the pellethane tubing was corrugated and torn.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the damage to the paddle is consistent with damage during use.
 
Event Description
This event is being reported based on the analysis of the returned device. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL (AFD-PLYMOUTH)
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key8961767
MDR Text Key156570585
Report Number3005334138-2019-00480
Device Sequence Number1
Product Code MTD
UDI-Device Identifier05415067028198
UDI-Public05415067028198
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/04/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2022
Device Model NumberD-AVHD-DF16
Device Lot Number6914405
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/27/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/21/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-