| Model Number |
GF-210RA |
| Medical Device Problem Codes |
Display or Visual Feedback Problem (1184); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The biomedical engineer reported that the multigas unit is intermittently giving an "external device failure" error message.The customer has tried using different cables, as well as replacing the sampling line, but the problem still persists.The customer was unable to confirm if the unit was in patient use at the time the issue occurred.The unit was sent in for further evaluation.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The biomedical engineer reported that the multigas unit is intermittently giving an "external device failure" error message.
|
| |
|
Event or Problem Description
|
|
The biomedical engineer reported that the multigas unit is intermittently giving an "external device failure" error message.
|
| |
|
Additional Manufacturer Narrative
|
|
Details of complaint: on (b)(6) 2019, customer stated gas unit intermittently experience "external device failure".Device was put into service on 3/2/2019.Service requested: exchange.Service performed: exchange.Investigation result: the investigation under (b)(4) concluded that for gf-210ra revision cb and onward with sensor unit cd-314p revision 2 (serial numbers: (b)(6) needed a firmware upgrade.Revision 2 of cd-314p utilize a new pump with a slight difference in specification.The sensor unit detects this small difference and output as gas device error.The resolution was to perform a firmware upgrade.For gf-210ra with older sensor unit cd-314p (no revision#: first revision), there was no tendency of sensor failing, thus the investigation provided no countermeasure.Due to no tendency of failure, it was determined these sensors failed as a result of normal usage.Cd-314p replacement is recommended.In summary, serial numbers: (b)(6) and below has the first version of cd-314p.Serial numbers: (b)(6).Have version 2 of cd-314p.These units require firmware upgrade.Serial numbers above (b)(6) are not affected.The root cause of the error message on serial number: (b)(6) was due to firmware requiring upgrade.Correction: d4.Unique device# (udi) : incorrect udi listed on mdr initial.Corrected on mdr follow up 001.Additional information: b4.Date of this report.F6.Date user facility/importer became aware of the event.F7.Type of report.F11.Date report sent to fda.F13.Date report sent to manufacturer.G4.Date received by manufacturer.G7.Type of report.H2.If follow-up, what type? additional information.Correction.Device evaluation.H3.Device evaluated by manufacturer? h6.Event problem and evaluation codes.H10.Additional manufacturer narrative.
|
| |
|