Product analysis: the product remains implanted; therefore, no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
|
Medtronic received additional information that 12 years and 4 months post implant of this aortic bioprosthetic valve, computed tomography (ct) angiogram revealed moderate restriction of the aortic valve and moderately thickened leaflets with significant pannus.At this time the internal diameter of the bioprosthesis was 14mm.No interventions were performed at that time.13 years and 6 months post implant, echocardiogram showed moderate to severe thickening and calcification of the leaflets, aortic stenosis, a peak gradient of 65mmhg and a mean gradient of 42mmhg.Ultimately 13 years and 7 months post implant, the valve was replaced with a transcatheter valve.No additional adverse patient effects were reported.Patient weight added in a4 relevant history added in b7 patient and device codes updated in h6.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|