ENCORE MEDICAL L.P. FMP HIP; FMP HEMISPHERICAL SHELLS W/SCREW HOLES, 48MM, W/P2 COATING
|
Back to Search Results |
|
| Model Number |
430-98-048 |
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
08/06/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reason for this revision surgery was reported as pain.The previous surgery and the surgery detailed in this investigation occurred 3.7 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to pain.There were no findings during this investigation that indicate that the reported device was defective.No other information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
|
| |
|
Event or Problem Description
|
|
Revision surgery - cup was never seated on original surgery, did not ingrow.Presented with pain post operation.Acetabular components were removed, larger size implanted.
|
| |
|
Search Alerts/Recalls
|
|
|