| Catalog Number |
45-27015 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/09/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Surgeon reported that during a surgery the tip of the screwdriver blade broke.Procedure was completed successfully with no adverse consequences or surgical delay reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed.Based on investigation, the root cause was attributed to be user related.The failure was caused by too high applied forces during use.The device inspection revealed the following: the tip of the screwdriver blade is indeed completely broken / snapped off.A close up of the fracture surface shows clearly some cracks, whereas the high applied forces led to a strain hardening / embrittlement of the material which finally resulted in the breakage.The root cause of the failure was repeated powerful dynamically overload during use (manufactured in april 2017).The ifu guide advises that during engaging the screw, axial pressure of the screwdriver into the screw head must be adequately applied to ensure that the blade is fully inserted into the screw head.This results in proper axial alignment and full contact between driver and screw, minimizing the risk of damage and leading to optimal friction fit performance.The lifetime of the device is not specified, but as stated in the cleaning and sterilization guide : ¿ [¿] stryker trauma & extremities typically does not specify the maximum number of uses appropriate for re-usable medical devices.The useful life of these devices depends on many factors including the method and duration of each use, and the handling between uses.Careful inspection and functional test of the device before use is the best method of determining the end of serviceable life for the medical device.[¿]¿.Based on the above-mentioned observations the case could be classified as user-related.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the investigation report will be updated.
|
| |
|
Event or Problem Description
|
|
Surgeon reported that during a surgery the tip of the screwdriver blade broke.Procedure was completed successfully with no adverse consequences or surgical delay reported.
|
| |
|