| Model Number |
MMT-7020B |
| Medical Device Problem Codes |
Incorrect, Inadequate or Imprecise Result or Readings (1535); Calibration Problem (2890)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the sensor had inaccurate readings that triggered threshold suspend alarm.The customer¿s blood glucose was 377 mg/dl.Sensor glucose level at the event was 118 mg/dl.The customer was informed that their blood glucose and sensor glucose levels were not in acceptable range 259 mg/dl.Insulin delivery was suspended due difference in blood and sensor glucose value.The sensor will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspected one opened and used sensor and found cannula bent and stuck to adhesive tape.Unable to confirm customer received sensor in said condition due to customer returned opened and used.
|
| |
|