Model Number N/A
Medical Device Problem Code Unraveled Material (1664)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/12/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.The actual device was not returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.A review of the device history record of the product code/lot# combination was conducted with no findings.
 
Event or Problem Description
The user facility reported that a physician was using a glidesheath slender nitinol kit and when traversing the wire in the sheath, he noticed something did not feel right.The physician pulled the wire out of the sheath to find it unraveled and unusable.The patient was in stable condition.The procedure outcome was successful.There was no patient injury or surgical intervention.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 to provide the completed investigation results.The evaluation of the actual device could not be conducted due to the device not being returned.With no device return, the exact cause of the reported event cannot be definitively determined based on the available information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESHEATH SLENDER
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
MDR Report Key8964788
Report Number1118880-2019-00239
Device Sequence Number2609955
Product Code DYB
UDI-Device Identifier00389701011363
UDI-Public00389701011363
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K173831
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2021
Device Model NumberN/A
Device Catalogue Number50-1050
Device Lot NumberXE17
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/14/2019
Supplement Date Received by Manufacturer10/10/2019
Initial Report FDA Received Date09/05/2019
Supplement Report FDA Received Date10/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-