MAUDE Adverse Event Report: BIOSENSE WEBSTER INC. WEBSTER ® CS CATHETER WITH EZ STEER® THECHNOLOGY AND AUTO ID; ELECTRODE RECORDING CATHETER OR ELECTRODE RECORDING PROBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC. WEBSTER ® CS CATHETER WITH EZ STEER® THECHNOLOGY AND AUTO ID; ELECTRODE RECORDING CATHETER OR ELECTRODE RECORDING PROBE Back to Search Results
Catalog Number BD710FJ282CT
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cardiac Tamponade (2226); No Code Available (3191)
Date of Event 08/12/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device investigation details: the product was discarded, therefore no product failure analysis can be conducted and device malfunction cannot be confirmed.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturing record evaluation cannot be conducted because no lot number was provided by the customer.(b)(4).Biosense webster manufacturer's reference number (b)(4) has two reports related to the same event: mfr # 2029046-2019-03598 for product code d134804 (thermocool® smart touch® sf bi-directional navigation catheter).(2) mfr # 2029046-2019-03599 for product code bd710fj282ct (webster ® cs catheter with ez steer® technology and auto id).
 
Event or Problem Description
It was reported that a female patient underwent an paroxysmal atrial fibrillation (afib) ablation procedure with a webster ® cs catheter with ez steer® technology and auto id and suffered cardiac tamponade requiring pericardiocentesis and surgical intervention.Post-ablation phase, cardiac tamponade was confirmed by echo as the patient¿s blood pressure dropped.Pericardiocentesis was performed to remove about 300-400 ml of fluid from the pericardial space.The patient was then moved to the operating room to close the perforation.During surgery it was found that there was a tear in the coronary sinus (cs).Patient¿s outcome is improved, the physician reported the patient was out of surgery and doing well.It is unknown if extended hospitalization was required as a result of the adverse event.Physician¿s opinion regarding the cause of the adverse event is that it was procedure-related, most likely due to either the cs catheter or the ablation catheter (which was located in the right atrium near the cs) perforating the cs during several multiple cardioversions.No bwi product malfunctions were reported.Transseptal puncture was performed with a st.Jude medical brk-1 xs transseptal needle.The flow was set at 15 ml/min.No error messages were observed on any bwi equipment.An 8.5 fr.St jude agilis sheath was used with the ablation catheter.The patient received anticoagulant (heparin) during the case with an activated clotting time (act) >350 seconds.The irrigation flow setting was set at 15 ml/min.The thermocool® smart touch® sf bi-directional navigation catheter was not in close proximity to another catheter.The thermocool® smart touch® sf bi-directional navigation catheter was zeroed upon initial warm up phase post connection to the carto 3 system and was not indicated to be re-zeroed.Since the physician believed the perforation may have occurred due to either the cs catheter or ablation catheter this event will be reported under both products.
 
Additional Manufacturer Narrative
On(b)(6)2019 , it was discovered that the thermocool® smart touch® sf bi-directional navigation catheter (product code d134804) was inadvertently omitted from the 3500a initial mdr that was submitted to fda on (b)(6)2019.The device has now been added to section d11.Concomitant products section.Manufacturer's ref # pc-000526106.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WEBSTER ® CS CATHETER WITH EZ STEER® THECHNOLOGY AND AUTO ID
Common Device Name
ELECTRODE RECORDING CATHETER OR ELECTRODE RECORDING PROBE
Manufacturer (Section D)
BIOSENSE WEBSTER INC.
33 technology drive
irvine CA 92618
MDR Report Key8964817
Report Number2029046-2019-03599
Device Sequence Number11415551
Product Code DRF
UDI-Device Identifier10846835002423
UDI-Public10846835002423
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K090898
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberBD710FJ282CT
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/05/2019
Supplement Date Received by Manufacturer09/05/2019
Initial Report FDA Received Date09/05/2019
Supplement Report FDA Received Date09/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
8.5 FR. ST JUDE AGILIS SHEATH; CARTO 3 SYSTEM; ST. JUDE BRK-1 XS NEEDLE; THMCL SMTCH SF BID, TC, F-J
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient Age65 YR
-
-