| Catalog Number |
TEXP135RUH |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
07/03/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Concomitant products: artoura breast tissue expander 700cc, catalog number: texp135ruh, serial number: (b)(4).Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) year-old caucasian female patient underwent a breast reconstruction primary with an artoura breast tissue expander 700cc and experienced deflation on the left side.As a result, the patient underwent removal on (b)(6) 2019.
|
| |
|
Additional Manufacturer Narrative
|
|
On 09/12/2019, the mentor failure analysis lab received the device for evaluation.Device evaluation summary: during evaluation of the device it was observed on anterior view four (4) tears, all tears measuring approximately 1 cm.Microscopic examination of the edges of each tear revealed parallel striations.No other anomalies were discovered.The striations are consistent with markings made by a sharp instrument perforating silicone material.As stated in the product insert data sheet, do not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps or scissors to contact the device during the implantation or other surgical procedures as this can result in rupture.It should be noted that as part of our quality process, each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|