|
|
| Catalog Number |
RBY2C1035 |
| Medical Device Problem Codes |
Physical Resistance/Sticking (4012); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/06/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report numbers: 3005168196-2019-01674, 3005168196-2019-01676.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the internal iliac artery using ruby coils and a lantern delivery microcatheter (lantern).During the procedure, the physician placed the lantern through a non-penumbra diagnostic catheter without issue.While advancing a ruby coil out of its introducer sheath and into the lantern, the physician experienced resistance.Subsequently, the ruby coil became stuck in the lantern and unintentionally detached when the physician attempted to pull it back; therefore, the ruby coil was removed by pulling the introducer sheath.The physician then advanced another ruby coil out of its introducer sheath and through the lantern.However, resistance was encountered at the tip of the lantern and when the physician attempted to retract the ruby coil, it unintentionally detached.Therefore, the lantern containing the detached ruby coil was removed.After removal, the physician noticed the lantern was kinked/ovalized three centimeters from the tip.The procedure was then completed using five other ruby coils, a new lantern, and the same diagnostic catheter.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Results: the pet lock was intact on the proximal end of its pusher assembly.The embolization coil was detached from its pusher assembly and the pull wire remained in its initial position within the distal detachment tip (ddt).The proximal constraint sphere was intact with the embolization coil.Conclusions: evaluation of the first returned ruby coil confirmed that the embolization coil was detached from its pusher assembly.If the ruby coil is forcefully retracted against resistance, the pusher assembly may elongate beyond the reach of its pull wire, allowing the embolization coil to detach.Further evaluation of the returned ruby coil revealed that the pusher assembly was kinked.These kinks were likely incidental to the reported complaint.Evaluation of the second returned ruby coil confirmed that the embolization coil was detached from its pusher assembly and was within the hub of the returned lantern.If the ruby coil is forcefully retracted against resistance, the pusher assembly may elongate beyond the reach of its pull wire, allowing the embolization coil to detach.Evaluation of the returned lantern revealed that the distal shaft of the catheter was kinked and ovalized.If the device is forcefully manipulated at extreme angles during use, damage such as these may occur.During functional testing, resistance was experienced while attempting to advance a demonstration ruby coil through the returned lantern due to the damage on the distal shaft.The kink and ovalization on the distal shaft likely contributed to the reported resistance experienced during manipulation of the both ruby coils within the lantern during the procedure.Penumbra coils and catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 1.3005168196-2019-01674.2.3005168196-2019-01676.
|
| |
|
Search Alerts/Recalls
|
|
|