|
The device was not returned for evaluation.A potential root cause could be the catheter is too stiff for insertion.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "warnings: on latex and red rubber catheters, do not use ointments or lubricants having a petrolatum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to failure, and/or to injury, illness or death of the patient.Note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Adverse reactions: urinary tract infection.Bleeding from the urethra.Irritation of the urethra.Instructions for intermittent catheters: wash your hands thoroughly with soap and water.Remove catheter from the pack.Position yourself comfortably, cleaning the opening of the urethra and surrounding area.Gently insert rounded end of catheter into urethra.When urine stops flowing, remove catheter from urethra.Dispose of catheter in accordance with local rules and regulations.Wash your hands.".
|