| Model Number |
MMT-723LNAS |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/28/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump's buttons were not working.The customer¿s blood glucose level was unknown at the time of the incident.Customer was advised to observe time clock for one minute to verify if time advances and they stated that working as suppose to.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with no button response on esc, act, up arrow and down arrow button due to flattened (no creased) dome switch and connector was inspected and unlocked on lcd board noted.The test reservoir p-cap was able to lock in place.
|
| |
|