| Model Number |
8253001RF |
| Medical Device Problem Code |
Failure to Sense (1559)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/14/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Other relevant device(s) are: product id: 8253200rf, serial/lot #: (b)(4), udi#: (b)(4).Analysis results were not available as of the date of this report. a follow-up report will be submitted when analysis is complete.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A health care provider (hcp) reported that nerve monitoring devices were not stimulating during the procedure.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the mainframe found that there was no fault in the device.The service and repair could not duplicate customer's complaint.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis of the mainframe is not yet available.Codes contained in this report is for the analysis of the patient interface with reference number below.Analysis confirms the alleged defect.The fuse was blown and the wave washers were worn.The item was tested and passed all manufacturing specifications.Product id: 8253200rf, serial/lot #: (b)(4), udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|