| Model Number |
DL-330-10-K |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/16/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Review of the rns neurostimulator ecogs and lead impedances confirmed that the data were suggestive of a potential lead break.The explanted lead was returned to neuropace for analysis.The investigation included visual inspection and electrical testing.Neuropace was unable to determine the cause of damage to the lead coil.
|
| |
|
Event or Problem Description
|
|
A review of the patient's ecog data indicated signal artifact on the right depth lead, suggestive of a possible lead break.The patient was brought into the clinic for reprogramming, and detection and stimulation were disabled on the affected lead.The patient denied any seizure-related falls; however the neurosurgeon noted it may be related to some type of trauma as the x-ray indicated a visible kink in the lead.The right depth lead was explanted on (b)(6) 2019 and a new depth lead was implanted and connected to the rns neurostimulator without complication.
|
| |
|