| Model Number |
6533 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 20mm x 3.00mm nc quantum apex balloon catheter was selected for treatment.However, it was noted that the balloon kinked and broke.No complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.:returned product consisted of an nc quantum apex balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 67.1cm distal of the strain relief.There was blood in the guidewire lumen.The balloon was tightly folded.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.A 20mm x 3.00mm nc quantum apex balloon catheter was selected for treatment.However, it was noted that the balloon kinked and broke.No complications were reported.
|
| |
|