| Model Number |
26265 |
| Medical Device Problem Codes |
Retraction Problem (1536); Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was received with the stent fully deployed.The device was received with the stent fully deployed from the device the device.It is not known when deployment occurred.The deployed stent was not returned for analysis.A visual and microscopic examination identified no damage to the stent cups or stent holder that could have contributed to the complaint incident.A visual and tactile examination found no damage or issues to the delivery system that could potentially have contributed to the complaint incident.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that stent partial deployment occurred and the device was removed without reconstraining.The 50-60% stenosed target lesion was located in the mildly tortuous and non-calcified transhepatic biliary.A 8x60x75//6f wallstent rp endoprosthesis was advanced for treatment.However, after the stent sheath reached the target position, the stent could not fully deploy.The stent was not fully reconstrained back into the catheter.The device was simply pulled back and removed.The procedure was completed with another of the same device.There were no patient complications reported and the patient's status was stable.
|
| |
|