| Catalog Number |
5530-G-513 |
| Medical Device Problem Codes |
Material Erosion (1214); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Injury (2348); Limited Mobility Of The Implanted Joint (2671)
|
| Date of Event |
07/09/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.Not returned.
|
| |
|
Event or Problem Description
|
|
Knee revision due to poly wear.
|
| |
|
Event or Problem Description
|
|
Knee revision due to poly wear.Update: additional comment provided by clinician: "review of x-rays reveal the tibial component is in 5 degrees in varus and 5 degrees in posterial slope.To more accurately determine alignment, full length x-rays would be required.".
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding wear involving a triathlon insert was reported.The event was confirmed based on evaluation of the returned device.Method & results product evaluation and results: material analysis concluded: the damage present on the posterior portion of the medial condyle was consistent with delamination.The damage present on the posterior portion of the lateral condyle was consistent with contact against the tibial baseplate.The damage on the articulating surface of the insert is consistent with burnishing, scratching, and third body indentations; which are common damage modes of uhmwpe.The yellow discoloration of the insert is consistent with absorption of synovial fluid.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Medical records received and evaluation: a review of the provided medical records by a clinical consultant stated the following comment: "review of x-rays reveal the tibial component is in 5 degrees in varus and 5 degrees in posterial slope.To more accurately determine alignment, full length x-rays would be required." product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the wear and delamination of the insert was confirmed based on the evaluation of the returned device.The exact cause of the event could not be determined because insufficient information was provided.Review of provided medical records reveal the tibial component is in 5 degrees in varus and 5 degrees in posterial slope which may have contributed to the wear of the device, however, the malposition cannot be confirmed as to more accurately determine alignment, full length x-rays would be required.No further investigation for this event is possible at this time.If additional information becomes available such as full length x-rays to indicate further evaluation is warranted, this record will be reopened.
|
| |
|