STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5530-G-513
Medical Device Problem Codes Material Erosion (1214); Naturally Worn (2988)
Health Effect - Clinical Codes Injury (2348); Limited Mobility Of The Implanted Joint (2671)
Date of Event 07/09/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.Not returned.
 
Event or Problem Description
Knee revision due to poly wear.
 
Event or Problem Description
Knee revision due to poly wear.Update: additional comment provided by clinician: "review of x-rays reveal the tibial component is in 5 degrees in varus and 5 degrees in posterial slope.To more accurately determine alignment, full length x-rays would be required.".
 
Additional Manufacturer Narrative
An event regarding wear involving a triathlon insert was reported.The event was confirmed based on evaluation of the returned device.Method & results product evaluation and results: material analysis concluded: the damage present on the posterior portion of the medial condyle was consistent with delamination.The damage present on the posterior portion of the lateral condyle was consistent with contact against the tibial baseplate.The damage on the articulating surface of the insert is consistent with burnishing, scratching, and third body indentations; which are common damage modes of uhmwpe.The yellow discoloration of the insert is consistent with absorption of synovial fluid.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Medical records received and evaluation: a review of the provided medical records by a clinical consultant stated the following comment: "review of x-rays reveal the tibial component is in 5 degrees in varus and 5 degrees in posterial slope.To more accurately determine alignment, full length x-rays would be required." product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the wear and delamination of the insert was confirmed based on the evaluation of the returned device.The exact cause of the event could not be determined because insufficient information was provided.Review of provided medical records reveal the tibial component is in 5 degrees in varus and 5 degrees in posterial slope which may have contributed to the wear of the device, however, the malposition cannot be confirmed as to more accurately determine alignment, full length x-rays would be required.No further investigation for this event is possible at this time.If additional information becomes available such as full length x-rays to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8975671
Report Number0002249697-2019-03142
Device Sequence Number18909770
Product Code MBH
UDI-Device Identifier07613327050554
UDI-Public07613327050554
Combination Product (Y/N)N
Initial Reporter CountrySE
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/31/2017
Device Catalogue Number5530-G-513
Device Lot NumberMLTJ3P
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2019
Initial Date Received by Manufacturer 08/13/2019
Supplement Date Received by Manufacturer11/29/2019
Initial Report FDA Received Date09/09/2019
Supplement Report FDA Received Date12/27/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age46 YR
Patient Weight86
-
-