COVIDIEN 760 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 760
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/19/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation summary: service engineer evaluated the ventilator and replaced the o2 (oxygen) pressure transducer.After servicing the ventilator passed all testing per manufacturing specification and was placed back into clinical use.If the replaced part is returned for failure investigation, a supplemental medwatch report with the findings will be submitted once the investigation is complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during use on a patient the 760 ventilator became inoperative with audible and visual alarm and stopped ventilation.It was also noted that ventilator failed post (power on self-test) after power cycling and has diagnostic error codes present in the logs.Patient was removed from the ventilator and placed on an alternate ventilator with no harm reported.
 
Additional Manufacturer Narrative
Additional codes added to section h6 evaluation code result and conclusion.Device evaluation summary: one oxygen pressure transducer was returned to medtronic for further analysis.The part was visually inspected and functionally tested with no anomalies or product deficiency identified.Although the field service engineer replaced the oxygen pressure transducer; however, the returned component was analyzed and tested satisfactorily.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
760 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway,gw
MDR Report Key8976110
Report Number8020893-2019-00179
Device Sequence Number2864242
Product Code CBK
UDI-Device Identifier10884521180970
UDI-Public10884521180970
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K990897
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number760
Device Catalogue NumberG-061949-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2019
Initial Date Received by Manufacturer 08/20/2019
Supplement Date Received by Manufacturer01/03/2020
Initial Report FDA Received Date09/09/2019
Supplement Report FDA Received Date01/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-