| Catalog Number |
0375542000 |
| Medical Device Problem Code |
Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/08/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that foreign material was found in sterile packaging.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: reported that there was red paint within the teeth of the burr the failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root cause/s could be it is possible that a small amount of plastic ¿red flash¿ was created during the inductive weld process and was not removed.The inner cutter accidentally scraped the housing hub during insertion and remained attached to the cutter teeth.Improper cleaning process, qc incoming, and in-process inspections.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that foreign material was found in sterile packaging.
|
| |
|
Search Alerts/Recalls
|