| Catalog Number |
CL10041001 |
| Medical Device Problem Code |
Thermal Decomposition of Device (1071)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/25/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Plant investigation: the actual device was returned to the manufacturer for physical evaluation.An external inspection of the received device found the usb device¿s strain relief damaged.There are no other discrepancies found externally.The crit-line iv monitor failed to recognize the clic device and the clic device¿s led did not illuminate.The monitor must recognize the device to perform a verify accuracy test or a field calibration.An internal inspection on the clic device found corrosion on the power supply board.There was evidence of an internal short on c104 and d102 of the power supply board.There is also corrosion on the receiver board and under the cover of the clic device.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.
|
| |
|
Event or Problem Description
|
|
A user facility reported that the clic device will not verify.After contacting the customer, it was confirmed that the replacement device was received.This issue was not an out of box failure either as it has been installed on the machine for a duration of time prior to this incident.There was no patient involvement associated with the reported event.Upon physical evaluation of the clic device, an internal short on c104 and c102 of the power supply board was identified.
|
| |
|
Search Alerts/Recalls
|