MAUDE Adverse Event Report: CONCORD MANUFACTURING CRIT-LINE CLIP (CLIC) - USB; ACCESSORIES, BLOOD CIRCUIT, HEMODIALYSIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING CRIT-LINE CLIP (CLIC) - USB; ACCESSORIES, BLOOD CIRCUIT, HEMODIALYSIS Back to Search Results
Catalog Number CL10041001
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/25/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Plant investigation: the actual device was returned to the manufacturer for physical evaluation.An external inspection of the received device found the usb device¿s strain relief damaged.There are no other discrepancies found externally.The crit-line iv monitor failed to recognize the clic device and the clic device¿s led did not illuminate.The monitor must recognize the device to perform a verify accuracy test or a field calibration.An internal inspection on the clic device found corrosion on the power supply board.There was evidence of an internal short on c104 and d102 of the power supply board.There is also corrosion on the receiver board and under the cover of the clic device.A review of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, a device history record (dhr) review was performed and verified that the results of the in-progress and final quality control (qc) testing met all requirements.
 
Event or Problem Description
A user facility reported that the clic device will not verify.After contacting the customer, it was confirmed that the replacement device was received.This issue was not an out of box failure either as it has been installed on the machine for a duration of time prior to this incident.There was no patient involvement associated with the reported event.Upon physical evaluation of the clic device, an internal short on c104 and c102 of the power supply board was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRIT-LINE CLIP (CLIC) - USB
Common Device Name
ACCESSORIES, BLOOD CIRCUIT, HEMODIALYSIS
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
MDR Report Key8976378
Report Number2937457-2019-02883
Device Sequence Number3295384
Product Code KOC
UDI-Device Identifier00840861101139
UDI-Public00840861101139
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K121599
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCL10041001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2019
Device Age MO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/29/2019
Initial Report FDA Received Date09/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-