• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® URETHRAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® URETHRAL CATHETER Back to Search Results
Model Number 802412
Medical Device Problem Code Misassembled (1398)
Health Effect - Clinical Codes Patient Problem/Medical Problem (2688); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the catheter had no eyelets, which restricted the flow.The patient alleged that he did not notice the device did not have eyelets until he inserted the device and no urine flowed, where he saw a small amount of blood.
 
Additional Manufacturer Narrative
The reported event was confirmed as manufacturing-related.Per evaluation, the red rubber intermittent catheter sample was examined and it was noted that the catheter was missing the drainage eyes.The root cause of this failure is operator error during the eye punching process due to which no eyes were punched, resulting in patient impact and no urine drainage.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿warnings: ¿ on latex and red rubber catheters, do not use ointments or lubricants having a petrolatum base.They will damage latex.¿ visually inspect the product for any imperfections or surface deterioration prior to use.¿ reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to failure, and/or to injury, illness or death of the patient.Note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Adverse reactions urinary tract infection bleeding from the urethra irritation of the urethra instructions for intermittent catheters: 1.Wash your hands thoroughly with soap and water.2.Remove catheter from the pack.3.Position yourself comfortably, cleaning the opening of the urethra and surrounding area.4.Gently insert rounded end of catheter into urethra.5.When urine stops flowing, remove catheter from urethra.6.Dispose of catheter in accordance with local rules and regulations.7.Wash your hands.".
 
Event or Problem Description
It was reported that the catheter had no eyelets, which restricted the flow.The patient alleged that he did not notice the device did not have eyelets until he inserted the device and no urine flowed, where he saw a small amount of blood.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDIA® URETHRAL CATHETER
Common Device Name
CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8976485
Report Number1018233-2019-05473
Device Sequence Number276137
Product Code KOD
UDI-Device Identifier00801741018343
UDI-Public(01)00801741018343
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,distributor,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/31/2022
Device Model Number802412
Device Catalogue Number802412
Device Lot NumberNGDP0530
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2019
Initial Date Received by Manufacturer 08/19/2019
Supplement Date Received by Manufacturer11/29/2019
Initial Report FDA Received Date09/09/2019
Supplement Report FDA Received Date12/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-