| Model Number |
977C165 |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received by a manufacturer representative (rep) regarding an implantable neurostimulator (ins) lead.It was reported that there was an impedance issue on the lead.The rep took the lead out of the battery multiple times, wiped the contacts down, re-inserted the lead and the same 2 electrodes kept showing impedance issues.The lead ports in the battery were switched and the same 1 electrode showed impedance issues.The lead was replaced and the issue was resolved.The lead was expected to be returned for analysis.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|