| Model Number |
TRUEMETRIX |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Codes |
Nausea (1970); Dizziness (2194)
|
| Date of Event |
08/14/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal report: (b)(4).Meter and test strips were not returned for evaluation.Most likely underlying root cause: mlc-30: user applied blood to strip before inserting strip into meter.Test strips udi: (b)(4).Note: manufacturer contacted customer in a follow-up call to ensure that the customer's condition improved - able to establish contact with customer and stated condition has improved.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for e-3 accompanied by symptoms.The expected fasting blood glucose test result range is undisclosed.The customer did report symptoms of feeling lightheaded and nauseous.Medical attention is not reported as a result of the actual blood glucose results or reported symptoms.The product storage location is undisclosed.During the call a blood test was performed by the customer and produced test results of 321mg/dl non-fasting using.The test strip lot manufacturer's expiration date is 12/29/2020 and open vial date is undisclosed.The meter memory was not reviewed for previous test result history.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for e-3 accompanied by symptoms.The expected fasting blood glucose test result range is undisclosed.The customer did report symptoms of feeling lightheaded and nauseous.Medical attention is not reported as a result of the actual blood glucose results or reported symptoms.The product storage location is undisclosed.During the call a blood test was performed by the customer and produced test results of 321mg/dl non-fasting using.The test strip lot manufacturer's expiration date is 12/29/2020 and open vial date is undisclosed.The meter memory was not reviewed for previous test result history.
|
| |
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) sections with additional information as of 15-may-2020: h6: updated fda's method, result, and conclusion codes h10: meter was not returned for evaluation.Test strips were not returned for evaluation.Retention testing was performed using test strips from the same lot.Retention strip lot tested and passed.Internal report: (b)(4).Meter and test strips were not returned for evaluation.Most likely underlying root cause: mlc-30: user applied blood to strip before inserting strip into meter.Test strips udi: (b)(4).Note: manufacturer contacted customer in a follow-up call to ensure that the customer's condition improved - able to establish contact with customer and stated condition has improved.
|
| |
|