• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE X; ULTRASONIC SURGICAL DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE X; ULTRASONIC SURGICAL DEVICE Back to Search Results
Model Number TB-0535FCX
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/08/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device was returned to olympus medical systems corp.(omsc) for evaluation.The probe was broken off at 12.5 mm from the distal end.There was scratch around the broken point and the coating for electric insulation of the probe was partially missing around the scratch.The area around the scratch showed that crack developed.The manufacturing record was reviewed and found no irregularities.Based on the past similar cases, it was known that the probe was cracked and the coating for electric insulation of the probe was damaged when the user activated output contacting with metal or something hard object such as metal clips, stapler, or other instruments, besides the probe was broken off when the probe was subjected to a load.The above device handling has warned in the instruction manual as follows.Do not grasp or let the probe tip contact hard objects such as metal clips, stapler, or other instruments (e.G., uterine manipulator).Also, be careful to avoid contacting the probe tip with those accidentally.Particularly during activation, a scratch on the probe tip could occur due to ultrasonic vibration, which leads the probe tip to break and fall off into the body cavity.In addition, the high-frequency (rf bipolar) current flows through the metal and generates spark discharge, which may cause burns and decrease functionalities.
 
Event or Problem Description
During a hepatobiliary and pancreatic procedure, the subject device was used.In procedure, the unspecified error occurred.The intended procedure was completed with another device.There was no patient injury reported.On (b)(6) 2019, olympus medical systems corp.(omsc) found that the probe was broken off and the coating for electric insulation of the probe was partially missing.This is the report regarding the breakage of the probe and the missing of the probe coating.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE X
Common Device Name
ULTRASONIC SURGICAL DEVICE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key8978563
Report Number8010047-2019-03232
Device Sequence Number11546429
Product Code GEI
UDI-Device Identifier04953170392207
UDI-Public04953170392207
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K180575
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberTB-0535FCX
Device Lot Number93K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2019
Initial Report FDA Received Date09/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-