| Model Number |
H700123 |
| Medical Device Problem Code |
Failure to Analyze Signal (1539)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/03/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The investigation results will be provided in a subsequent submission. .
|
| |
|
Event or Problem Description
|
|
During a diagnostic exploration procedure, the signal leads would not display on the recording system and the procedure was cancelled.The procedure was completed at a different facility.There were no adverse consequences to the patient due to the cancellation.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record was not possible since the system serial number was provided and the component serial number was not.The results of the investigation are inconclusive since the device was not returned for analysis.Additional information from the field confirmed the system is functional and still in use at the customer facility.Based on the information received, the cause of the reported signal loss and subsequent procedure cancellation could not be conclusively determined.
|
| |
|
Search Alerts/Recalls
|