• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. MAKO X3 UNI ONLAY TIBIAL INSERT SIZE 4 - 8 MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. MAKO X3 UNI ONLAY TIBIAL INSERT SIZE 4 - 8 MM; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Catalog Number 180734-1
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Pain (1994); Injury (2348)
Date of Event 08/14/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An event regarding revision due to pain involving a mako insert was reported.The event was not confirmed.Method & results: device evaluation and results: the device was not returned.Visual inspection of the received image of the explant noted no discrepancy on the insert.Dimensional, functional and material analysis could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Device history review: review of the product history records indicates devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there has been no other event for the lot referenced.Conclusions: it was reported that the patient's left knee was converted from pka to tka due to pain.Visual inspection of the received image of the explant noted no discrepancy on the insert.The event could not be confirmed nor the root cause be determined as insufficient information was available.No further investigation for this event is possible at this time as no devices and / or insufficient information was received by stryker orthopaedics.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.The following devices were also listed in this report: mck femoral-lm-rl-sz 3; cat# 180503; lot# 508346-m.Mck tibial baseplate-lm/rl-sz 4; cat# 180604; lot# 26221016-01.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
 
Event or Problem Description
It was reported that the patient's left knee was converted from a pka to a tka due to pain (possible cause or contributor to pain unknown).A mako pka was converted to a triathlon tka.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAKO X3 UNI ONLAY TIBIAL INSERT SIZE 4 - 8 MM
Common Device Name
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key8980990
Report Number3005985723-2019-00658
Device Sequence Number365038
Product Code NPJ
UDI-Device Identifier00848486016494
UDI-Public00848486016494
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K150307
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Catalogue Number180734-1
Device Lot NumberR474R9
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/14/2019
Initial Report FDA Received Date09/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age55 YR
Patient Weight110
-
-