MAUDE Adverse Event Report: MEDTRONIC IRELAND ENDURANT II ILIAC STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ENDURANT II ILIAC STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number ETLW1610C124E
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Injury (2348)
Date of Event 07/16/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An endurant stent graft system was implanted in a patient for the endovascular treatment of a 44 mm diameter abdominal aortic aneurysm.The infra-renal neck was small in diameter measuring 14 mm and long measuring 27 mm.Approximately four years later the physician reported that the neck has dilated to 55 mm.It was noted that a non mdt stent that was placed with the endurant bifurcate in the index procedure is not opposed to the stent graft and there is a new type ia endoleak.An intervention was carried out to treat the type ia endoleak with heli-fx endoanchors which were placed posterior and to the right where the endoleak was visualized.It was noted that the endoanchors were not successful in resolving the endoleak.It was also noted during the intervention procedure that a type ib endoleak ws observed at the right side and a non mdt stent graft was added to extend the limb by 15 mm, landing proximal to the riia.This endoleak was resolved.No additional clinical sequelae were reported and the patient will be monitored.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the type ib endoleak appeared to be coming from the iliac dilating around the vessel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II ILIAC STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key8981083
Report Number9612164-2019-03839
Device Sequence Number19678276
Product Code MIH
UDI-Device Identifier00643169467644
UDI-Public00643169467644
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/27/2017
Device Model NumberETLW1610C124E
Device Catalogue NumberETLW1610C124E
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/16/2019
Supplement Date Received by Manufacturer09/10/2019
Initial Report FDA Received Date09/10/2019
Supplement Report FDA Received Date10/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-