| Model Number |
367004 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The pantera leo balloon catheter was selected for use.During the second predilatation, the balloon ruptured at 20 bar.Therefore, the device was removed and another balloon catheter was used.
|
| |
|
Additional Manufacturer Narrative
|
|
Neither the complaint instrument nor the angiographic material was provided for analysis.Therefore no technical investigation on the subject could be performed.The manufacturing history was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test.The instrument was delivered in a leak-proof condition.Based on the conducted investigations no manufacturing related root cause could be determined.The root cause is most likely related to external factors during procedure.
|
| |
|