Model Number 801764
Device Problem Failure to Run on Battery (1466)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/10/2019
Event Type  malfunction  
Manufacturer Narrative
Per the manufacturer's subsidiary, the battery icon color on the central control monitor (ccm) and the light emitting diode (led) on the front panel of the hlm were both green.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the heart lung machine (hlm) would not switch from alternating current (ac) to battery power.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
Evaluation is in progress, but not yet concluded.
 
Manufacturer Narrative
Per the manufacturer's subsidiary, the power manager board was defective.
 
Manufacturer Narrative
Updated blocks: h3 and h6 the reported complaint was confirmed.During laboratory analysis, the product surveillance technician observed the power manager could not switch from alternating current to battery.Microscopic examination of the power manager boards found that there were several charred resistors.The resistance across r186, r159 and r122 resistors showed that the circuits were open.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key8982861
MDR Text Key157147108
Report Number1828100-2019-00480
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup,Followup
Report Date 04/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801764
Device Catalogue Number801764
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/20/2019
Initial Date FDA Received09/10/2019
Supplement Dates Manufacturer Received10/22/2019
12/11/2019
04/06/2020
Supplement Dates FDA Received11/14/2019
12/18/2019
04/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-