No patient information provided as no patient was involved in this concern.A medtronic representative went to the site to test the equipment.Testing revealed that the power cable for the navigation system was replaced to restore functionality.The system then passed the system checkout and was found to be fully functional.The power cable was returned to the manufacturer for evaluation.Testing found that the power cable had an open in a brown wire in the cable.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.Other relevant device(s) are: product id: 9733678, serial/lot #: (b)(4).If information is provided in the future, a supplemental report will be issued.
|