| Model Number |
400CE |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A customer report was received on (b)(6) 2019 (cdr # (b)(4)) stating the heomtherm heater-cooler alarmed with an ee03 code.It was determined the thermistor required replacement.
|
| |
|
Event or Problem Description
|
|
A customer reported the hemotherm 400ce went down during a case with an ee03 error.No patient injury was reported.
|
| |
|
Search Alerts/Recalls
|