• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC NAVIGATION STEALTH STATION S7 SYSTEM; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC NAVIGATION STEALTH STATION S7 SYSTEM; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9733856
Device Problem Human-Device Interface Problem (2949)
Patient Problems No Known Impact Or Consequence To Patient (2692); Radiation Exposure, Unintended (3164)
Event Date 08/15/2019
Event Type  malfunction  
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that the issue could not be replicated.The system then passed the system checkout and was found to be fully functional.Device manufacturing date is unavailable.Other relevant device(s) are: 9733686 synergy spine software.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information regarding a navigation device being used intra/peri-operatively of a sacroiliac and thoracolumbar procedure.It was reported that the site's second spin had failed to transfer.The site took one spin, navigated without issue, and when they took the second spin, it wouldn't automatically transfer.They were unable to manually transfer either.The site was able to complete the procedure with the first spin and the 2nd spin was not required.There was a less than 1-hour delay to the procedure and no impact on patient outcome.
 
Manufacturer Narrative
The logs for the navigation system were reviewed by medtronic personnel.However, the logs provided no additional insight into the probable cause of the anomaly.The software investigation found that a probable cause was unable to be determined without further information since the on-going investigation proved to be inconclusive based on the information provided.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIGATION STEALTH STATION S7 SYSTEM
Type of Device
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key8983879
MDR Text Key157363757
Report Number1723170-2019-04831
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9733856
Device Catalogue Number9733856
Was Device Available for Evaluation? No
Date Manufacturer Received10/14/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age61 YR
Patient Weight112
-
-