| Model Number |
MMT-332A |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
09/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the reservoir was leak at o ring's and insulin pump received insulin flow blocked alarm.The customer¿s blood glucose level was unknown.The customer stated that insulin pump was exposed to moisture and insulin was leaked to the reservoir compartment.It was also reported that customer declined troubleshooting for high blood glucose and insulin flow blocked alarm.The customer stated that they did not received any alarms immediately after moisture exposure and insulin pump did not passed self test.The reservoir will be returned and insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluated one open/used reservoir.Inspected reservoir's o-rings/stopper groove for anomalies, none were found.Ran basal, bolus leaking test : reservoir filled with diluent and connected to a new infusion set.Installed reservoir and connector into a paradigm insulin pump.Conclusion: reservoir do not leak and not occluded.(b)(4).
|
| |
|