| Model Number |
MMT-723NAP |
| Medical Device Problem Codes |
Mechanical Problem (1384); Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/31/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that their insulin pump had a motor error.The customer¿s blood glucose was 127 mg/dl.Customer reported that the insulin pump had no delivery alarm.Customer wanted to replace the insulin pump.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test.No excessive no delivery/occlusion alarm noted.No motor error alarm noted.Motor passed motor test.
|
| |
|