• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION SIMULUS FULLY FLEXIBLE ANNULOPLASTY RING; RING, ANNULOPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION SIMULUS FULLY FLEXIBLE ANNULOPLASTY RING; RING, ANNULOPLASTY Back to Search Results
Model Number 700FF35
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes No Information (3190); No Code Available (3191)
Date of Event 08/14/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that immediately post implant of this 35mm mitral annuloplasty band, it was explanted and replaced with a 39mm annuloplasty band of the same model.The reason for replacement was not reported.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Medtronic received additional information that the 35mm band was too small for the patient's annulus, subsequently resulting in systolic anterior motion of the mitral valve.There was no allegation against the 35mm band itself.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMULUS FULLY FLEXIBLE ANNULOPLASTY RING
Common Device Name
RING, ANNULOPLASTY
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key8992648
Report Number2025587-2019-02785
Device Sequence Number3380954
Product Code KRH
UDI-Device Identifier00643169182523
UDI-Public00643169182523
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K052565
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/02/2024
Device Model Number700FF35
Device Catalogue Number700FF35
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/01/2019
Supplement Date Received by Manufacturer09/30/2019
Initial Report FDA Received Date09/11/2019
Supplement Report FDA Received Date10/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient Weight64
-
-