• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL LRS PLATELET, PLASMA, RBC SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL LRS PLATELET, PLASMA, RBC SET Back to Search Results
Catalog Number 82400
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 08/19/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation: the device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.The trima accel procedure summary (taps) report was submitted to aid in the investigation.The report confirms there were no flags indicating that the platelet product should be verified for white blood cell contamination.The run data file (rdf) was analyzed for this event.Root cause: a definitive root cause for the observed leukoreduction failure remains undetermined at this time.The signals in the run data file indicate that the higher-than-expected wbc content in the platelet product was likely a result of an escape of wbc from the lrs chamber during a portion of the procedure.Based on the available information, it cannot be ruled out that the higher-than expected wbc content in the platelet product could be donor-related.
 
Event or Problem Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.There was not a transfusion recipient or patient involved at the time of the residual white blood cell (rwbc) testing, therefore no patient information is reasonably known at the time of the event.Donor unit (b)(6).The platelet collection set is not available for return because it was discarded by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Common Device Name
TRIMA ACCEL LRS PLATELET, PLASMA, RBC SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
MDR Report Key8992809
Report Number1722028-2019-00260
Device Sequence Number14183559
Product Code GKT
UDI-Device Identifier05020583824003
UDI-Public05020583824003
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
BK170059
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2020
Device Catalogue Number82400
Device Lot Number1803225151
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/20/2019
Initial Report FDA Received Date09/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-