| Model Number |
85231499 |
| Medical Device Problem Codes |
Particulates (1451); Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); Not Applicable (3189)
|
| Date of Event |
08/15/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Bayer product analysis received and examined the returned low pressure connector tubing (lpct).Visual examination confirmed the presence of a particle within the tube set.The particle was identified as degraded resin from the tubing extrusion process and was partially connected to the inside wall of the tubing.Our investigation determined that the particle noted in the lpct was introduced during the manufacturing process and was not detected upon receipt of the product from the supplier.A 12 month review of complaint history determined the complaint rate to be 0.30 complaints per million (cpm).
|
| |
|
Event or Problem Description
|
|
The customer reported the following: after opening the stellant ct disposable kit the customer observed a foreign body inside the tubing.No injury or adverse event was reported.
|
| |
|