| Model Number |
5540 |
| Medical Device Problem Code |
Separation Problem (4043)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Langston v2 catheter was returned for evaluation.A manufacturing record review was completed and zero nonconformances were found therefore supporting that the device met material, assembly and performance specifications.Catheter was sent out to the supplier for investigation.When information is received a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
Catheter split near the hub after power injection.900psi 24 for 12.No patient injury or impact reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Supplier attempted to recreate the defect to determine if excessive glue may have been the root cause of the catheter rupture.Thirty test units were created and tested per current testing processes for burst pressure.At this time, between the visual examination of the returned unit and the testing of replicate units, we are unable to establish a definite root cause.If further information is received a follow-up report will be submitted.
|
| |
|