| Catalog Number |
B15LT |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).See facility medwatch number (b)(4).Additional information was requested and the following information was received: how was the procedure completed? torn off piece was retrieved and procedure completed in the normal manner.(b)(4).
|
| |
|
Event or Problem Description
|
|
See user facility medwatch number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # t93932.Investigation summary the analysis results found that the b15lt device was received with no apparent damage.In addition, a piece plastic was returned inside of a plastic bag.Further analysis the piece of plastic belongs to the duckbill.The tyvek was also returned along with the instrument.A leak test was performed, and the device was noted to leak without the test probe inserted through the device.No conclusion could be reached as to what might have caused the reported event.The reported complaint was confirmed.A manufacturing record evaluation was performed for the finished device batch number and no non-conformances were identified.
|
| |
|