• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT ALUMINA INSERT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT ALUMINA INSERT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE Back to Search Results
Catalog Number 625-0T-36G
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code Injury (2348)
Date of Event 08/14/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Device not available.
 
Event or Problem Description
The surgeon reported that he undertook a revision of a trident exeter hip.**update** the customer reported that the exeter trunnion was worn.The ceramic head was broken and the liner had wear marks.
 
Additional Manufacturer Narrative
An event regarding wear involving a trident liner was reported.The event was confirmed.Method & results: device evaluation and results: visual inspection was performed as part of the material evaluation and indicated the following comments: metal transfer markings observed on distal side of liner.Mar required.Visual inspection was performed as part of the material analysis report (mar), dated 18-september-2019.This inspection indicated: damage consistent with impingement against the stem was observed on rim of the insert and sleeve.Metal transfer markings and scratches were observed on the articulating surface of the alumina insert.Damage consistent with the explantation process was observed on the proximal surface of the sleeve.Clinician review: a review of the provided medical records and/or x-rays by a clinical consultant indicated: 8/13/19 brief handwritten operative note.".Post.Approach.Metalosis lymph nodes and 100 ml graphite black.Excision/lavaged.Multiple ceramic head fragments/stem/ceramic liner removed.Cement mantle intact.New components inserted." no clinical or past medical history, no comprehensive operative reports, no x-rays, no examination of explanted components, and no surgical pathology reports.While the brief, handwritten revision operative note seems to partially confirm the event description, there is insufficient data to create a report for this event.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, and x-rays are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Event or Problem Description
The surgeon reported that he undertook a revision of a trident exeter hip.The customer reported that the exeter trunnion was worn.The ceramic head was broken and the liner had wear marks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT ALUMINA INSERT
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8997522
Report Number0002249697-2019-03184
Device Sequence Number10854020
Product Code MRA
UDI-Device Identifier04546540516862
UDI-Public04546540516862
Combination Product (Y/N)N
Initial Reporter CountryGB
PMA/510(K) Number
P000013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Catalogue Number625-0T-36G
Device Lot Number60862703
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2019
Initial Date Received by Manufacturer 08/16/2019
Supplement Date Received by Manufacturer10/25/2019
Initial Report FDA Received Date09/11/2019
Supplement Report FDA Received Date11/22/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age62 YR
Patient Weight140
-
-